AI ScaleKnowledge

Weighing Scale Error Codes

Find immediate solutions for system faults, calibration errors, and maintenance alerts across all supported models.

STABLE ERROR

Stable Measurement Not Achieved

STABLE ERROR appears when the balance cannot obtain a stable reading within the expected time. Possible Causes β€’ Air movement β€’ Vibration β€’ Moving sample β€’ Electromagnetic interference Troubleshooting 1. Close all draft shield doors. 2. Place the balance on a stable table. 3. Reduce airflow. 4. Wait for the stability indicator. 5. Repeat the measurement. Stable environmental conditions improve weighing performance.

Read Fix β†’
MEMORY ERROR

Internal Memory Failure

MEMORY ERROR indicates that stored balance parameters cannot be accessed correctly. Possible Causes β€’ Memory corruption β€’ Power interruption β€’ Internal hardware fault Troubleshooting 1. Restart the balance. 2. Restore factory settings if available. 3. Reconfigure operating parameters. 4. Perform calibration. If MEMORY ERROR continues, professional service is recommended.

Read Fix β†’
REPEAT ERROR

Repeatability Test Failed

REPEAT ERROR indicates that consecutive measurements do not meet repeatability requirements. Possible Causes β€’ Drafts β€’ Vibration β€’ Temperature fluctuations β€’ Unstable installation Troubleshooting 1. Close draft shield doors. 2. Verify proper leveling. 3. Eliminate vibration sources. 4. Perform response adjustment. 5. Repeat the measurement. Stable environmental conditions are essential for analytical balance repeatability.

Read Fix β†’
OVERLOAD

Maximum Capacity Exceeded

OVERLOAD appears when the applied weight exceeds the maximum weighing capacity. Possible Causes β€’ Excessive sample weight β€’ Mechanical overload β€’ Foreign object under weighing pan Troubleshooting 1. Remove excess weight immediately. 2. Inspect the weighing pan. 3. Restart the balance. 4. Verify normal zero indication. Preventing overload protects the precision weighing sensor from permanent damage.

Read Fix β†’
EC ERROR

Electronic Circuit Error

EC ERROR indicates that the balance has detected an abnormal condition within its electronic circuitry. Possible Causes β€’ Power interruption β€’ Internal PCB fault β€’ Voltage instability β€’ Hardware malfunction Troubleshooting 1. Switch OFF the balance. 2. Disconnect power for one minute. 3. Restart the balance. 4. Perform internal calibration. 5. Verify stable power supply. If EC ERROR remains, contact authorized A&D service personnel.

Read Fix β†’
LO

Sample Too Light

LO indicates that the sample weight is below the minimum required value for the selected function. Possible Causes β€’ Sample too small β€’ Counting reference weight insufficient β€’ Percent reference weight too low Troubleshooting 1. Increase sample quantity. 2. Use a heavier reference sample. 3. Verify weighing mode. 4. Check balance sensitivity. The LO message disappears once a valid reference weight is detected.

Read Fix β†’
CH NG

Response Adjustment Failed

CH NG appears when automatic response adjustment cannot determine a stable weighing condition. Possible Causes β€’ Excessive vibration β€’ Strong air movement β€’ Unstable table β€’ Continuous movement on weighing pan Troubleshooting 1. Move the balance to a stable location. 2. Close all weighing chamber doors. 3. Remove vibration sources. 4. Restart automatic response adjustment. 5. Verify proper leveling. Stable environmental conditions are required for successful response adjustment.

Read Fix β†’
CAL ERROR

Calibration Failed

CAL ERROR indicates that the calibration process could not be completed successfully. Possible Causes β€’ Vibration during calibration β€’ Air drafts affecting measurement β€’ Unstable power supply β€’ Internal calibration failure β€’ Incorrect external calibration weight Troubleshooting 1. Remove all objects from the weighing pan. 2. Restart the balance. 3. Allow one hour warm-up time. 4. Place the balance on a stable surface. 5. Repeat calibration. If CAL ERROR continues after multiple attempts, internal service or inspection may be required.

Read Fix β†’
INTERNAL ERROR

Internal System Error

INTERNAL ERROR indicates that the balance has detected an internal hardware or software fault during operation. Possible Causes β€’ Temporary system malfunction. β€’ Memory initialization failure. β€’ Electronic circuit fault. β€’ Power interruption during startup. β€’ Internal component failure. Troubleshooting Steps 1. Switch OFF the balance. 2. Disconnect the power adapter for one minute. 3. Restart the balance. 4. Allow the balance to complete initialization. 5. Perform calibration and verify normal operation. 6. If the error remains, discontinue use and contact an authorized A&D service center. Preventive Tips β€’ Use only the original A&D power adapter. β€’ Avoid sudden power interruptions. β€’ Keep the balance away from moisture and excessive heat. β€’ Schedule periodic preventive maintenance.

Read Fix β†’
STABILITY ERROR

Unstable Weight Reading

STABILITY ERROR occurs when the balance cannot obtain a stable weighing value. Possible Causes β€’ Air movement around the balance. β€’ Table vibration. β€’ Improper leveling. β€’ Unstable power supply. β€’ Draft shield left open. β€’ Electromagnetic interference. Troubleshooting Steps 1. Place the balance on a solid and vibration-free surface. 2. Close all draft shields. 3. Check that the leveling bubble is centered. 4. Remove nearby fans or air conditioning airflow. 5. Wait until the stability indicator appears before recording weight. 6. Restart the balance if the problem continues. Preventive Tips β€’ Install the balance in a temperature-controlled environment. β€’ Calibrate the balance regularly. β€’ Keep the weighing chamber clean.

Read Fix β†’
UNDERLOAD

Negative Load Error

UNDERLOAD indicates that the balance is detecting a value below its normal zero reference. Possible Causes β€’ Weighing pan is not installed correctly. β€’ Load cell is under tension. β€’ Balance was powered ON without the weighing pan. β€’ Mechanical obstruction under the pan. β€’ Internal sensor issue. Troubleshooting Steps 1. Switch OFF the balance. 2. Verify that the weighing pan and pan support are installed correctly. 3. Remove any foreign objects touching the weighing mechanism. 4. Restart the balance and allow it to complete initialization. 5. Perform zero calibration if required. 6. If UNDERLOAD continues to appear, inspect the load cell assembly and contact authorized A&D service. Preventive Tips β€’ Always install the weighing pan before powering ON the balance. β€’ Keep the weighing chamber clean. β€’ Avoid pulling upward on the weighing pan.

Read Fix β†’
OVERLOAD

Over Capacity Error

Possible Causes β€’ Load exceeds maximum capacity β€’ Object stuck under pan β€’ Mechanical obstruction Solutions 1. Remove load. 2. Inspect weighing pan. 3. Restart balance. 4. Perform calibration. 5. Contact service if required.

Read Fix β†’
CAL ERROR

Calibration Failed

Possible Causes β€’ Incorrect calibration weight β€’ Unstable environment β€’ Vibration β€’ Air draft β€’ Wrong calibration procedure Troubleshooting 1. Restart the balance. 2. Verify calibration weight. 3. Check leveling. 4. Remove vibration. 5. Repeat calibration. Professional service may be required if the error persists.

Read Fix β†’
Lo

Low Battery Voltage

Model Compatibility: A&D Portable Balances and Battery-Equipped Models Solution: The internal battery or external power supply voltage has dropped below the operational threshold. 1.) Action: If operating on batteries, replace all batteries with fresh ones immediately. 2.) Action: If using an AC adapter, ensure it is the original A&D power supply specifically designed for this model. 3.) Warning: Using incorrect voltage or amperage will trigger this fault or potentially damage the internal power board. 4.) Check: Inspect the power cable for any signs of fraying or damage.

Read Fix β†’
CAL E

Calibration Error

Model Compatibility: A&D Analytical and Precision Balances (GF, GX, HR, FX series) Solution: The balance detected an invalid calibration mass or excessive environmental interference during the calibration process. 1.) Action: Verify that you are using the exact weight value requested by the balance display (e.g., 100.000g). 2.) Check: Ensure your calibration weight is highly accurate (Class E2 or F1) and is not damaged or severely oxidized. 3.) Procedure: Perform the calibration strictly in a draft-free and vibration-free environment. 4.) Procedure: Ensure the pan is completely empty before pressing the PRINT/ENTER key at the 'CAL 0' step.

Read Fix β†’
Error 1

Stability Error

Model Compatibility: A&D Analytical and Precision Balances (GF, GX, HR, FX series) Solution: The balance cannot stabilize enough to display a reading or complete a function (like taring or calibration). 1.) Action: Close all breeze break doors securely. 2.) Environment: Check for strong drafts from air conditioners, open windows, or cooling fans. 3.) Environment: Ensure the weighing table is completely isolated from vibrations caused by nearby machinery or foot traffic. 4.) Setting: If issues persist, enter the function settings and adjust the environmental response settings to "Slow" or "Stable" mode.

Read Fix β†’
-E or Under

Underload / Missing Pan Error

Model Compatibility: A&D Analytical and Precision Balances (GF, GX, HR, FX series) Solution: The load on the sensor is too light, falling below the zero point range. 1.) Action: Ensure the weighing pan and pan support are properly installed on the balance. 2.) Action: Press the [RE-ZERO] or [TARE] key to attempt to reset the zero point. 3.) Check: Ensure the balance is not tilted and that nothing is lifting the pan upward from underneath the scale.

Read Fix β†’
E or Over

Overload Error

Model Compatibility: A&D Analytical and Precision Balances (GF, GX, HR, FX series) Solution: The weight placed on the pan exceeds the maximum weighing capacity of the balance. 1.) Action: Immediately remove the object from the weighing pan to prevent damage to the load cell. 2.) Check: Inspect the pan and surroundings for any debris or physical interference pushing down on it. Note: If the error persists with an empty pan, the load cell may have suffered permanent mechanical shock or requires a master zero recalibration.

Read Fix β†’
Err 08

Err 08 - Tare/Zero Command Failed

βš™οΈ Model Compatibility: Aczet CG, CZ, and Precision Series Balances. "Err 08" appears when the operator presses the [TARE] or [ZERO] key, but the scale cannot execute the command. Troubleshooting Steps: 1. Instability: The scale will not tare if the weight reading is fluctuating. Ensure the draft shield is closed and the table is free from vibrations. 2. Capacity Limit: You cannot tare a container that exceeds the maximum capacity of the balance. Use a lighter weighing boat or container. 3. Negative Value: If the scale was already tared with a heavy weight, and that weight is removed (leaving a large negative number), pressing zero may trigger this error. Turn the scale OFF and ON to clear the negative memory.

Read Fix β†’
Err 08

Err 08 - Tare/Zero Command Failed

βš™οΈ Model Compatibility: Aczet CG, CZ, and Precision Series Balances. "Err 08" appears when the operator presses the [TARE] or [ZERO] key, but the scale cannot execute the command. Troubleshooting Steps: 1. Instability: The scale will not tare if the weight reading is fluctuating. Ensure the draft shield is closed and the table is free from vibrations. 2. Capacity Limit: You cannot tare a container that exceeds the maximum capacity of the balance. Use a lighter weighing boat or container. 3. Negative Value: If the scale was already tared with a heavy weight, and that weight is removed (leaving a large negative number), pressing zero may trigger this error. Turn the scale OFF and ON to clear the negative memory.

Read Fix β†’
Err 01 - Reference Weight Too Low

Err 01 - Reference Weight Too Low

βš™οΈ Model Compatibility: Aczet CG, CZ, and Precision Series Balances. "Err 01" triggers specifically when you are using the "Piece Counting" or "Percentage Weighing" applications. It means the sample weight you placed on the pan to establish the initial 100% reference (or the 10-piece reference) is too light for the load cell to calculate an accurate baseline. How to Fix: 1. Increase your sample size. Instead of placing 10 pieces on the scale to establish the reference weight, place 50 or 100 pieces. 2. Once the scale has a heavier physical baseline to work with, it will accept the reference, and you can begin counting or percentage weighing normally.

Read Fix β†’
Login Failed

Login Failed - Authentication Security

βš™οΈ Model Compatibility: Aczet CM Series (21 CFR Compliant). A "Login Failed" message indicates that the entered username or password does not match the system records. Troubleshooting Steps: 1. Ensure the correct "User Type" (Administrator, IT Admin, User, or Master) is selected before entering credentials. 2. Password Sensitivity: Passwords are case-sensitive. Check if the Caps Lock is active. 3. User Disabled: If you are an administrator, verify the user status in the "User ID Management" menu to see if the account has been "Disabled" by an admin. 4. Account Lockout: Repeated failed attempts may trigger a temporary lockout for security reasons. Wait for the specified lockout period or have an Administrator reset the account.

Read Fix β†’
Audit Trail Error

Audit Trail Error - Memory/Compliance Failure

βš™οΈ Model Compatibility: Aczet CM Series (21 CFR Compliant). The "Audit Trail Error" indicates that the internal memory reserved for 21 CFR compliance logging is either full or experiencing a write-error. Troubleshooting Steps: 1. Backup/Export Data: Navigate to the Audit Report menu and export all logs via USB or RS-232 to ensure compliance data is preserved. 2. Clear Memory: After a successful backup, use the administrative settings to clear the audit logs to free up memory space. 3. System Integrity: If the error persists, it may indicate a corrupt database file. Perform a backup of all User IDs and settings before attempting a factory reset.

Read Fix β†’
Error 2

Error 2 - Startup Load Error

βš™οΈ Model Compatibility: Aczet CY-C Series. "Error 2" occurs during the power-on self-test. It means the balance has detected an invalid weight (load) on the weighing pan while trying to initialize its zero-point. Troubleshooting: 1. Ensure the weighing pan is completely empty and correctly seated on the shaft. 2. Power the balance OFF and ON. If the error persists, it may indicate a mechanical issue with the load cell.

Read Fix β†’
--OL--

--OL-- (Overload) - Capacity Exceeded

βš™οΈ Model Compatibility: Aczet CY-C Series. The "--OL--" message indicates that the weight on the pan has exceeded the balance's maximum rated capacity (e.g., 220g). Immediate Action: 1. Immediately remove the item from the weighing pan to prevent damage to the high-precision load cell. 2. If "--OL--" remains visible while the pan is completely empty, it indicates a catastrophic load cell failure, likely caused by severe physical shock. Please contact an authorized Aczet service center.

Read Fix β†’
Error 2

Error 2 - Startup Load Error (CY-C Series)

βš™οΈ Model Compatibility: Aczet CY-C Series Analytical & Precision Balances. "Error 2" occurs during the power-on self-test. It means the balance has detected an invalid weight (load) on the weighing pan while it is trying to initialize its internal zero-point. Troubleshooting Steps: 1. Ensure the weighing pan is completely empty and that no foreign objects are touching it. 2. Check that the weighing pan is correctly seated on the weighing shaft. If it is tilted or touching the draft shield, the balance will misread the start-up load. 3. Power the balance OFF and then back ON. 4. If the error persists after ensuring the pan is clean and correctly seated, the balance may require a zero-point re-calibration by a qualified technician.

Read Fix β†’
Err 4

Err 4 - Span Calibration Failure

βš™οΈ Model Compatibility: Essae DS-852 & DS-852G. "Err 4" occurs if the span calibration process failed because the scale could not register a stable, correct reading during the test weight phase. Troubleshooting Steps: 1. Verify your test weight is certified and clean (any dust or oil on the test weight will cause a failure). 2. Ensure the test weight is placed precisely in the center of the platter. 3. Ensure there is no air movement in the roomβ€”high-precision scales for gold will fail calibration if the scale detects even a slight breeze during the "stabilization" phase.

Read Fix β†’
Lo bAt

Lo bAt - Low Battery Warning

βš™οΈ Model Compatibility: Essae DS-415 & DS-415N. The "Lo bAt" indicator warns that the internal rechargeable battery is near exhaustion. Action Required: 1. Connect the scale to a 230V AC mains outlet using the original Essae power adapter. 2. Leave the scale on charge for at least 8-10 hours. 3. If the scale does not hold a charge after a full cycle, the internal battery has reached its end-of-life and requires replacement.

Read Fix β†’
EEEE / O-L

EEEE / O-L - Maximum Capacity Exceeded

βš™οΈ Model Compatibility: Essae DS-415 & DS-415N. The "EEEE" or "O-L" (Overload) error indicates that the item weight exceeds the scale's maximum rated capacity (e.g., 30kg). Immediate Troubleshooting: 1. Remove the load from the platter immediately to prevent long-term damage to the internal load cell. 2. Power the scale OFF and then ON again. If the error remains while the platter is empty, it indicates mechanical stress on the load cell. This may require a re-calibration or, in worst-case scenarios, a load cell replacement.

Read Fix β†’
UndEr

UndEr or Dashes (- - - -) - Negative Weight Error

βš™οΈ Model Compatibility: Essae DS-252. The "UndEr" error or a row of negative dashes indicates that the scale is reading a weight significantly below its initial calibrated zero point. Troubleshooting Steps: 1. Ensure the stainless steel platter is actually placed on the scale. Operating the scale without the top platter will trigger this error. 2. Check beneath the platter to ensure no debris or packing material is pushing upward against the spider cross bracket. 3. Press the [REZERO] key to establish a new zero point. If the error refuses to clear, the load cell wiring may be damaged.

Read Fix β†’
Lo bAt

Lo bAt / Flashing Display - Low Battery Warning

βš™οΈ Model Compatibility: Essae DS-252. If the display begins to flash or shows the message "Lo bAt", the internal 6V/4Ah rechargeable battery is nearly depleted. How to Fix: 1. Plug the power cord into a stable 230V AC outlet. 2. Allow the scale to charge continuously for 8 hours. The DS-252 can be used normally while charging. 3. If the scale immediately shuts off when the AC cord is removed after a full charge cycle, the internal lead-acid battery has reached the end of its lifespan and must be replaced.

Read Fix β†’
EEEE

EEEE - Maximum Capacity Exceeded on DS-252

βš™οΈ Model Compatibility: Essae DS-252 (Bench and Table Top). The "EEEE" or "OvrFLw" error indicates that the weight placed on the platter exceeds the scale's maximum programmed capacity (e.g., 30kg or 50kg) plus the allowed 9 divisions. Immediate Troubleshooting Steps: 1. Immediately remove the heavy load from the platter to protect the aluminum load cell from permanent deformation. 2. Turn the machine OFF and then back ON using the power switch. Hardware Diagnostics: If the screen displays "EEEE" continuously while the platter is completely empty, the load cell has been permanently damaged by physical shock or extreme overloading. A technician will need to replace the load cell and recalibrate the machine.

Read Fix β†’
No Disk

No Disk or File Error - USB Dumping Failure

βš™οΈ Model Compatibility: Essae SI-850N System Indicator. These errors occur during data extraction or report printing: 1. "No Disk": The USB Pendrive was disconnected abruptly from the USB Host port before the data dump was completed. Ensure the flash drive is formatted to FAT16 or FAT32 (NTFS is not supported) and seated firmly. 2. "File Error": The scale is attempting to print or dump a "Custom Format" report, but the required custom format files have not been uploaded to the scale from a PC. Switch to a "Standard" format or upload the custom files.

Read Fix β†’
Unit Wt < 0.01%

Unit Wt < 0.01% of Accuracy! - Piece Weight Too Light

βš™οΈ Model Compatibility: Essae SI-850N System Indicator. When sampling parts to establish a base unit weight, this error indicates the individual piece is physically too light for the load cell's current resolution configuration. How to Fix: 1. Increase the sample size drastically (e.g., sample 50 or 100 pieces instead of 10) so the scale can register a stable baseline weight. 2. Go to Menu > Counting Settings > Unit Weight Average, and increase the setting to 64 for maximum sampling precision. 3. If the error persists, the physical items are too light for this specific scale's capacity, and a lower-capacity, higher-resolution platform is required.

Read Fix β†’
Accumulated

Accumulated or Quantity Error - Batching Limitations

βš™οΈ Model Compatibility: Essae SI-850N System Indicator. When using the Manual Accumulation feature (pressing the 'Acumulat' soft key), operators may encounter two specific warnings: 1. "Accumulated": This message appears if the operator presses the accumulate key a second time for the exact same batch without first removing the weight from the platter. The scale prevents double-counting. Simply remove the item, let the scale return to zero, and place the next item. 2. "Quantity Error": This triggers in the accumulation window if the total accumulated quantity exceeds 999,999 pieces. At this point, the batch must be closed or the counters must be reset to continue logging.

Read Fix β†’
Disp Resolution Err

Disp Resolution Err - Display Resolution Limit Exceeded

βš™οΈ Model Compatibility: Essae SI-850N System Indicator. The "Disp Resolution Err" triggers during the power-on self-check sequence if the basic capacity divided by the basic accuracy exceeds 15,000 divisions. How to Fix: This is a configuration issue, not a broken load cell. A technician must enter the Scale Menu and adjust the Basic Weight and Accuracy parameters so the total division ratio remains at or below 15,000.

Read Fix β†’
Wt Not Ready

Wt Not Ready - Invalid Logging Condition

βš™οΈ Model Compatibility: Essae SI-850N System Indicator. The "Wt Not Ready" message appears when an operator attempts to log a transaction, but the weight on the platform is either exactly zero or registering as a negative value. Troubleshooting Steps: 1. The SI-850 will strictly refuse to log zero or negative weights to prevent corrupt database entries. 2. Ensure the item is fully resting on the platform. 3. If the scale reads a negative number when empty, press the [Rezero] key. If a tare weight was left active, press the [Tare] key to clear it.

Read Fix β†’
Memory Low

Memory Low - Internal Data Memory Almost Full

βš™οΈ Model Compatibility: Essae SI-850N System Indicator. The "Memory Low" error triggers when the available internal data memory drops below 25 KB. The SI-850 stores up to 25,000 transaction records, and this warning means it is running out of capacity to log new weights. How to Fix: 1. Connect a FAT-formatted USB Pen Drive to the USB Host port. 2. The display will show "Pendrive: Connected". 3. Press the [Store] soft key to back up your records to a .LOG file. 4. Once the dump says "Printed Successfully", go to Menu > Log Settings > Delete Logged Records. 5. Press [Yes] to clear the internal memory.

Read Fix β†’
Load Cell Dead

Unresponsive Load Cell - Welding or Shock Damage

βš™οΈ Hardware Compatibility: Essae TWS Shear Beam Load Cell Kits[cite: 11]. If the load cell is completely unresponsive immediately after a new tank installation, it was likely destroyed by electrical currents or high heat during the fabrication process. Fabrication Rules for TWS Kits: 1. Never leave the load cell inside the kit during top or bottom plate welding[cite: 11]. 2. Before welding, un-screw the two Allen bolts, carefully remove the load cell along with the steel ball, and store it safely away from the welding area[cite: 11]. 3. Ensure the temperature of the fabrication structure coming into contact with the load cell kit remains below 45Β°C[cite: 11]. 4. If your load cell has been shorted out by welding currents, it cannot be repaired. You will need to purchase a replacement shear beam load cell.

Read Fix β†’
Erratic Tank Weight

Erratic Weight Fluctuations - Mechanical Friction or Vibration

βš™οΈ Hardware Compatibility: Essae TWS Load Cell Kits (Up to 10t)[cite: 11]. If your digital indicator (like the DX-451 or SI-850) is displaying fluctuating or inaccurate weights, the tank is likely experiencing mechanical binding or external vibrations[cite: 11]. Immediate Troubleshooting Steps: 1. Check the Piping: All inlet and outlet pipes attached to the tank must use flexible hoses[cite: 11]. Never use rigid pipes, as they transfer weight away from the load cells and cause massive weighing errors[cite: 11]. 2. Halt the Stirrer: If the tank uses an internal stirrer, it must be turned completely off[cite: 11]. Wait for the liquid to settle and take the weight reading only under stable conditions[cite: 11]. 3. Install Anti-Vibration Pads: If the environment is prone to vibration, you must install anti-vibration pads between the tank leg and the top plate, as well as below the bottom plate[cite: 11]. Note: Adding these pads will increase the total height of the load cell kit by approximately 30mm[cite: 11]. 4. Cable Routing: Ensure that the load cell cables are routed completely away from AC power lines to prevent electrical interference[cite: 11]. Never cross or route them alongside AC lines[cite: 11].

Read Fix β†’
Total Price > Rs. 99999.9

Total Price > Rs. 99999.9 - Transaction Limit Exceeded

βš™οΈ Model Compatibility: Essae SI-810PR Label Scale[cite: 10]. This message is triggered during a transaction if the calculated total price exceeds the maximum display limit of 99999.9[cite: 10]. How to Fix: You must reduce the quantity or weight of the items being billed in a single transaction to stay within the allowable limit[cite: 10]. Split large bulk orders into multiple smaller bills.

Read Fix β†’
Stock NOT there!

Stock NOT there! - Inventory Limit Reached

βš™οΈ Model Compatibility: Essae SI-810PR Label Scale[cite: 10]. The "Stock NOT there!" error occurs if internal inventory tracking is enabled and the available PLU stock is less than the item weight or quantity you are trying to transact[cite: 10]. How to Fix: A store manager must update the PLU stock levels[cite: 10]. 1. Enter Menu > Programming > Inventory[cite: 10]. 2. Enter the PLU Number and press the [PLU] key[cite: 10]. 3. Enter the newly arrived stock value and press the [Stock] soft key to add it to the existing inventory[cite: 10].

Read Fix β†’
PRINTER HEAD OPEN! / NO PAPER!

PRINTER HEAD OPEN! or NO PAPER! - Printer Hardware Errors

βš™οΈ Model Compatibility: Essae SI-810PR Label Scale[cite: 10]. In fast-paced retail environments, swapping label rolls quickly can lead to these common printer errors: 1. "PRINTER HEAD OPEN!": This message appears when the thermal print head is not closed down completely[cite: 10]. 2. "PRINTER DOOR OPEN!": This indicates that the main printer door is not fully shut[cite: 10]. 3. "NO PAPER!": This triggers when no print media is present in the machine[cite: 10]. 4. "TAKE PRINTED LABEL!": The scale will halt operations if the last printed label is not removed from its current position[cite: 10]. Troubleshooting: Open the printer door and verify that the label liner passes correctly through the Label Gap Sensor, between the platen and print head, and into the re-winder slit[cite: 10].

Read Fix β†’
Modbus Error 04

Modbus Exception Code 04 - Invalid Data Read

βš™οΈ Model Compatibility: Essae SI-810 System Indicator. When integrating the SI-810 into a factory PLC using RS-485 Modbus ASCII, the scale will return an Exception Code "04" (Invalid Data Read) if the scale itself is experiencing a physical weighing error[cite: 9]. What is causing it: This exception code is generated when the master queries the scale, but the scale cannot provide a valid weight string because it is currently in one of the following hardware error states[cite: 9]: 1. Weight Underflow (Load cell reading below zero)[cite: 9]. 2. Weight Overflow (Capacity exceeded)[cite: 9]. 3. Disp Resolution Error (Capacity/Accuracy ratio misconfigured)[cite: 9]. Troubleshooting for Integrators: Do not troubleshoot your communication protocol. An Exception 04 means your parity, baud rate, and LRC checksums are perfectly fine, but the physical scale platter is obstructed or overloaded. Have an operator clear the physical scale platform and send the read request (Function Code 04) again[cite: 9].

Read Fix β†’
Disp Resoln.

Disp Resoln. - Display Resolution Limit Exceeded

βš™οΈ Model Compatibility: Essae SI-810 System Indicator[cite: 9]. The "Disp Resoln." error triggers during the power-on diagnostic sequence or when configuring scale parameters[cite: 9]. It indicates that the basic capacity divided by the basic accuracy exceeds 15,000 divisions[cite: 9]. How to Fix: This is a software configuration issue. A technician must enter the Scale menu and adjust the Basic Weight and Accuracy parameters to ensure the total division ratio remains at or below 15,000 (e.g., 15kg capacity by 1g accuracy)[cite: 9].

Read Fix β†’
Wt Not Ready

Wt Not Ready - Invalid Logging Condition

βš™οΈ Model Compatibility: Essae SI-810 System Indicator[cite: 9]. The "Wt Not Ready" message appears in the weight window or accumulation window when an operator attempts to log a transaction, but the weight on the platform is either zero or registering as a negative value[cite: 9]. Troubleshooting Steps: 1. Ensure the item is fully resting on the platform and the scale has stabilized[cite: 9]. 2. Check if a Tare weight was accidentally left active from a previous batch. Press the [Rezero] or [Tare] key to clear the offset before placing the new item[cite: 9]. 3. Zero and negative weights cannot be logged into the system[cite: 9].

Read Fix β†’
Memory Low

Memory Low - Internal Data Memory Almost Full

βš™οΈ Model Compatibility: Essae SI-810 System Indicator[cite: 9]. The "Memory Low" error occurs when the available internal data memory drops below 25 KB[cite: 9]. The SI-810 is designed to hold up to 25,000 logged records, and this warning means you are about to run out of space[cite: 9]. How to Fix: 1. You must back up your data and clear the memory. Connect a FAT32-formatted USB Pen Drive to the USB Host port[cite: 9]. 2. The display will show "Pendrive: Connected" and generate a file name (e.g., 12300612.LOG)[cite: 9]. 3. Press the [Store] soft key to dump the records[cite: 9]. 4. Once the dump is complete, navigate to Menu > Log Settings > Delete Logged Records and press [Yes] to clear the internal memory[cite: 9].

Read Fix β†’
Disp Resolution

Disp Resolution - Invalid Capacity Configuration

βš™οΈ Model Compatibility: Essae RX-850 & RX-8XX Series Label Scales[cite: 8]. The "Disp Resolution" error triggers during the power-on diagnostic sequence[cite: 8]. It means the scale's internal capacity divided by its accuracy setting exceeds 15,500 divisions, which is outside the machine's legal metrology limits[cite: 8]. How to Fix: A technician must enter the Scale Configuration menu and adjust the Basic Weight parameters[cite: 8]. For example, setting a 15kg scale to a 1g accuracy is 15,000 divisions (Valid). Setting it to 0.5g would exceed the 15,500 limit and cause this error.

Read Fix β†’
Stock NOT there!

Stock NOT there! - PLU Inventory Limit Reached

βš™οΈ Model Compatibility: Essae RX-850 & RX-8XX Series Label Scales[cite: 8]. The "Stock NOT there!" error occurs when your scale's internal inventory tracking is enabled (Menu > Settings > Inventory Enable = Yes), and the available stock for the recalled PLU is less than the weight or quantity you are currently trying to bill[cite: 8]. How to Fix: A store manager must update the PLU stock levels[cite: 8]. 1. Enter Menu > Programming > Inventory[cite: 8]. 2. Enter the PLU Number and press the [PLU] key[cite: 8]. 3. Enter the newly arrived stock value and press the [Stock] soft key to add it to the existing inventory[cite: 8].

Read Fix β†’
PRINTER HEAD OPEN

PRINTER HEAD OPEN! or NO PAPER! - Printer Hardware Error

βš™οΈ Model Compatibility: Essae RX-850 & RX-8XX Series Label Scales[cite: 8]. In a fast-paced retail environment, changing label rolls quickly often leads to these two common errors: 1. "PRINTER HEAD OPEN!": This message indicates that the thermal print head mechanism is not latched down completely[cite: 8]. Press down firmly on both sides of the printer door until you hear it click into place. If the door is completely open, it will display "PRINTER DOOR OPEN!" instead[cite: 8]. 2. "NO PAPER!": This triggers when the label roll is empty or the internal gap sensor isn't detecting the backing paper[cite: 8]. Troubleshooting: Open the printer door and ensure the label liner passes cleanly through the Label Gap Sensor, between the platen and head, and into the re-winder slit[cite: 8]. Always remove the last printed label, or the machine will halt and display "TAKE PRINTED LABEL!"[cite: 8].

Read Fix β†’
Set Point Programming

How to Program Set Points on Essae DS-450 & DS-451

βš™οΈ Model Compatibility: Essae DS-450, DS-450SS, DS-451, and DS-451 Flame Proof[cite: 6]. These scales support up to 5 TTL outputs or 2 potential-free relay contacts for automation[cite: 6]. To program the Set Points for these outputs: 1. Ensure the scale is in normal weighing mode showing 0.000[cite: 6]. 2. Depress the [REZERO] key and press the [↑] (Digit Increment) key three times in sequence[cite: 6]. 3. The display will toggle between the message "SetPt-1" and the current programmed value[cite: 6]. 4. Use the [←] (Digit Select) and [↑] (Digit Increment) keys to enter your required target value[cite: 6]. 5. Press the [*] (Star) key to store the value entered and advance to the next set point (e.g., SetPt-2)[cite: 6]. 6. Repeat the entry process up to SetPt 5[cite: 6]. 7. Press the [T] (Tare) key to exit the Set Point function and return to weighing mode[cite: 6].

Read Fix β†’
shutoff

shutoff / Battery Flashing - Deep Discharge

βš™οΈ Model Compatibility: Essae DS-450, DS-450SS, and DS-451 (Note: Internal batteries are generally not applicable for Flame Proof machines)[cite: 6]. If the battery indicator is flashing or the display outright shows "shutoff", it indicates a low battery or deep discharge state[cite: 6]. Troubleshooting Steps: 1. Connect the machine to an active 230V AC power supply[cite: 6]. 2. Recharge the battery continuously for 8 to 10 hours[cite: 6]. 3. If the scale immediately shuts off when AC power is removed after a full charge cycle, replace the battery[cite: 6].

Read Fix β†’
Under flow

Under flow - Platter or Load Cell Issue

βš™οΈ Model Compatibility: Essae DS-450, DS-450SS, DS-451, and DS-451 Flame Proof[cite: 6]. The "Under flow" error happens when there is an obstruction or disconnection under the weighing pan, causing the scale to read below its calibrated zero point[cite: 6]. Troubleshooting Steps: 1. Check if any foreign material is touching the platform and pushing against the load cell[cite: 6]. Remove any foreign material found[cite: 6]. 2. Ensure the machine platter is not lifted and fix the platter firmly in its designated position[cite: 6]. 3. Verify that the load cell connection is not discarded, loose, or severed[cite: 6]. Check the cable of the load cell and connect it properly[cite: 6].

Read Fix β†’
Over flow

Over flow - Capacity Exceeded on DS-450/451

βš™οΈ Model Compatibility: Essae DS-450, DS-450SS, DS-451, and DS-451 Flame Proof[cite: 6]. The "Over flow" error indicates that the weighed material exceeds the maximum capacity of the machine[cite: 6]. Immediate Troubleshooting Steps: 1. Check the loaded capacity and immediately remove the extra weight from the platter to protect the load cell[cite: 6]. 2. Power cycle the scale. Hardware Diagnostics: If the scale displays "Over flow" when the platter is completely empty, the internal load cell has likely sustained severe shock damage and requires replacement.

Read Fix β†’
PF rEt

PF rEt Screen Message - Power Fail Retain Activation

βš™οΈ Model Compatibility: Essae DS-215HD Heavy Duty Scale. If your scale briefly displays "PF rEt" when turning on, this is not a fault. It stands for "Power Fail Retain." What it means: If the AC power supply is suddenly cut off while a heavy load is sitting on the platter, the scale remembers the exact weight. When power is restored, it displays "PF rEt" and immediately recovers the active weight reading so you do not lose your transaction. Conditions for this to work: 1. The weight on the scale must be greater than 0.5% of the machine's total full-scale capacity. 2. The internal Customer Specification parameter "T13" must have its MSD set to 1.

Read Fix β†’
MEMORY NOT CONNECTED

MEMORY NOT CONNECTED - Data Logging & Printer Error

βš™οΈ Model Compatibility: Essae DS-215HD Heavy Duty Scale. If your scale displays "MEMORY NOT CONNECTED" when trying to save a transaction or print a receipt, the motherboard is failing to communicate with the internal 32KB data memory chip, or data saving has not been properly enabled. Troubleshooting Steps: 1. Verify that your scale is set to "Custom Print Format". The DS-215HD can only save data memory records when using a custom format, not standard formats. 2. Check the hardware specification T15. (MSD must be set properly to allow both saving and printing). 3. To clear a corrupted memory bank, you can connect a PC via RS-232 and send the command "Ctrl+D" to delete all records. Hardware Note: If the internal EEPROM memory chip has burned out due to electrical surges, the scale will no longer store your 1000 transaction records. If memory tracking is critical to your dispatch operations, contact UP Scales via WhatsApp to upgrade to a modern, fully networked weighing terminal.

Read Fix β†’
shutoff

shutoff Display / Battery Indicator Flashing - Low Power

βš™οΈ Model Compatibility: Essae DC-85N & DC-851N Counting Scales. If the display shows "shutoff" or the battery indicator begins flashing, the internal 7.4V/2.2AH rechargeable Lithium-Ion battery has entered a deep discharge state or low battery indication[cite: 103, 371]. Immediate Troubleshooting Steps: 1. Connect the scale to an active 230V AC mains supply[cite: 103]. 2. Recharge the battery continuously for 8 to 10 hours. Hardware Diagnostics: If the machine displays "shutoff" the moment you disconnect the AC power cord, the internal battery has reached the end of its life cycle and must be replaced.

Read Fix β†’
Under flow

Under flow - Platter Disconnected or Obstructed

βš™οΈ Model Compatibility: Essae DC-85N & DC-851N Counting Scales. The "Under flow" error triggers when the scale registers a baseline weight significantly lower than its calibrated zero point. Immediate Troubleshooting Steps: 1. Ensure that the stainless steel or mild steel platter is firmly and correctly fixed onto the scale. 2. Check beneath the platform to remove any foreign material or debris that may be touching the platform and pushing it upward. 3. Check the load cell cable connection to ensure it has not been discarded or pinched. If the error persists after clearing the platter and checking connections, the load cell sensor may be malfunctioning. Reach out to the UP Scales service team for repair parts or a machine upgrade.

Read Fix β†’
Over flow

Over flow - Maximum Weighing Capacity Exceeded

βš™οΈ Model Compatibility: Essae DC-85N & DC-851N Counting Scales. The "Over flow" error indicates that the material placed on the scale exceeds the maximum rated capacity of the machine. Immediate Troubleshooting Steps: 1. Check the loaded capacity and immediately remove the extra weight from the platter to protect the load cell from permanent strain. 2. Power cycle the scale by turning it OFF and ON. Hardware Diagnostics & Sales Note: If the scale displays "Over flow" when the platter is completely empty, the internal load cell has likely sustained severe shock damage and requires replacement. In many cases, upgrading to a newer, higher-capacity industrial scale is more reliable than replacing heavily abused internal components. Contact UP Scales via WhatsApp to get a quote on a rugged, high-capacity replacement today!

Read Fix β†’
Disp Resolution Err

Disp Resolution Err - Invalid Capacity Configuration

βš™οΈ Model Compatibility: Essae DC-810 & DC-815 Label Scales. The "Disp Resolution Err" message is triggered during the Power-On sequence if the scale's internal capacity and accuracy settings have been programmed incorrectly. Specifically, it occurs if the calculated ratio of (Capacity Γ· Accuracy) exceeds 15,000 divisions. How to Fix This: 1. This is a configuration error, not a hardware failure. 2. You must enter the Settings Menu and correct the Basic Capacity and Accuracy parameters. (For example, a 15kg scale with a 1g accuracy equals 15,000 divisions, which is valid. Setting it to 0.5g would exceed the limit and cause this error). 3. If you are locked out of the menu, you must use the master override password to access the hardware configuration.

Read Fix β†’
Unit Wt < 0.01%

Unit Wt < 0.01% of Accuracy! - Piece Weight Too Light

βš™οΈ Model Compatibility: Essae DC-810 & DC-815 Counting Scales. When sampling parts to find the unit weight, the "Unit Wt < 0.01% of Accuracy!" error occurs. This means the individual piece you are trying to count is physically too light for the specific load cell capacity of your current DC-810 configuration. Troubleshooting Steps: 1. Increase your sample size. Instead of sampling 10 pieces, press the [PCS] key and manually enter a larger sample batch (e.g., 50 or 100 pieces) so the scale can register a stable baseline weight. 2. Check your "Unit Weight Average" setting under Menu > Counting Settings. Ensure it is set between 16 and 64 for maximum precision. Sales & Metrology Note: If you frequently need to count extremely lightweight components (like tiny electronic resistors or micro-screws), your current scale does not have a fine enough resolution. You need a dedicated High-Resolution Counting Scale or a Dual-Platform setup. Connect with UP Scales on WhatsApp to find the exact high-precision scale required for your micro-components!

Read Fix β†’
Weight Overflow

Weight Overflow - Capacity Exceeded on DC-810

βš™οΈ Model Compatibility: Essae DC-810 and DC-815 Counting Scales. The "Weight Overflow" (or simply "Overflow" in check weighing mode) message indicates that the load placed on the platter exceeds the machine's maximum rated capacity plus 9 divisions (Capacity + 9d). Immediate Troubleshooting Steps: 1. Immediately remove the heavy load from the platter to prevent permanent strain on the aluminum load cell. 2. Power cycle the scale using the ON/OFF rocker switch. Hardware Diagnostics: If the scale displays "Weight Overflow" continuously when the platter is completely empty, the internal load cell or the Analog-to-Digital (ADC) conversion circuitry has sustained permanent mechanical shock damage. Sales & Upgrade Note: Industrial environments are tough on precision counting scales. If your load cell is permanently damaged, repairing a heavily used DC-810 might result in recurring downtime. Explore our catalog of rugged, high-capacity industrial counting scales or contact the UP Scales sales team on WhatsApp for a replacement quote.

Read Fix β†’
Frozen Weight Display

Weight Value Field Frozen / No Change in Display Output

βš™οΈ Model Compatibility: Essae AX Series Balances. This system symptom occurs when placing a known mass or sample onto the platter results in zero variation or movement on the digital LED screen. Root Causes: - Improper initial platter positioning or binding of the inner core mechanism. - Mechanical locking pins or transit components are obstructing internal travel. Corrective Diagnostics & Countermeasures: 1. Remove the stainless steel pan completely and carefully re-seat it square onto the pan pins. 2. Inspect carefully inside the weighing area through the glass shield to ensure the metallic pan structure is floating independently and not touching the external housing outer framework.

Read Fix β†’
No Stable Indication

Missing Stable Indicator / Fluctuating and Floating Weight Readings

βš™οΈ Model Compatibility: Essae AX Series Precision Balances. When the "Stable" lamp (the small circle/indicator on the lower left of the LED display panel) fails to illuminate, or the values continue to drift constantly, the balance is struggling to isolate its high-resolution sensor from environment factors. Root Causes: - The instrument is resting on an unanchored or vibrating workspace table. - Severe air drafts or air conditioning crosswinds passing across the open draft shield glass doors. Corrective Diagnostics & Countermeasures: 1. Relocate the balance onto a heavy, completely solid anti-vibration lab bench surface. 2. Close all sliding side and top doors of the protective glass draft shield completely during measurement cycles. 3. Enter the customer programming options and increase your digital filter dampening values by setting parameter "C19 LSD" to 3 (High) or 4 (Very High).

Read Fix β†’
SHUTOF

SHUTOF System Message - Critical Low Battery Protection Shutdown

βš™οΈ Model Compatibility: Essae AX Balances equipped with Lithium-Ion backup options. The "SHUTOF" text sequence appears as a final safety warning immediately before the balance powers down its internal processor. This occurs when the alternative 7.4V / 2.2Ah battery bank drains to an unstable voltage floor. Corrective Diagnostics & Countermeasures: 1. Immediately connect the factory-supplied 9.5V DC external power adaptor to the rear DC inlet port. 2. Confirm the main wall outlet switch is active. 3. Let the unit remain plugged in continuously for roughly 6 hours to execute a safe battery replenishment cycle. Sales Note: If your balance returns to the "SHUTOF" state immediately after disconnecting the power cable, the internal battery pack has deteriorated completely. Consider upgrading to a brand-new model from the UP Scales catalog today.

Read Fix β†’
Und Flo

Und Flo Error Message - Underload Status / Missing Platter Assembly

βš™οΈ Model Compatibility: Essae AX Series Balances. The "Und Flo" code triggers when the system detects an internal measurement baseline that falls significantly below the calibrated factory zero tracking limit. Root Causes: - The upper stainless steel weighing platter has been removed or is improperly seated. - Debris or materials have wedged underneath the core support column, forcing it upward. Corrective Diagnostics & Countermeasures: 1. Ensure the platter assembly is present and placed perfectly centered onto its structural support slots. 2. Thoroughly check beneath the weighing pan area to confirm no foreign objects are causing upward tension on the column. 3. Once properly assembled, tap the [REZERO] button on the mechanical tact keyboard to re-index the zero value point.

Read Fix β†’
Ovr Flo

Ovr Flo Error Message - Platter Capacity Overload Limit Exceeded

βš™οΈ Model Compatibility: Essae AX Series Balances. The "Ovr Flo" display code signifies that a weight greater than the absolute maximum rated threshold of the scale has been placed onto the weighing pan. (For instance, over 220 grams on an AX-220 model). Root Causes: - Excess physical material resting directly on the pan. - Shock loading or someone manually pressing downward on the sensitive scale platform. - Permanent mechanical deformation of the Single Block Sensor system. Corrective Diagnostics & Countermeasures: 1. Immediately clear all materials, weights, and containers off the stainless steel platter. 2. Inspect around the edges of the pan to guarantee no physical components are jammed or pressing down on the support mechanism. 3. Power cycle the machine off for 10 seconds and turn it back on. Sales & Upgrade Advisory: If your balance continues to lock onto "Ovr Flo" while completely empty, the high-precision magnetic force core sensor has suffered terminal strain damage. Replacing delicate sensor blocks can often equal the cost of an entirely new device. To avoid extensive downtime, use the links below to shop our new range of laboratory balances or speak directly with an upgrade advisor on WhatsApp.

Read Fix β†’
ERROR 82

A/D Conversion Error

ERROR 82 indicates an abnormal analog-to-digital conversion condition. Possible Causes: β€’ A/D circuit failure β€’ Load cell wiring issue β€’ Electrical interference β€’ Damaged indicator PCB Troubleshooting Steps: 1. Switch OFF the indicator. 2. Inspect load cell wiring. 3. Verify grounding connections. 4. Check excitation voltage. 5. Restart the indicator. If ERROR 82 remains, professional service may be required.

Read Fix β†’
ERROR 15

Span Calibration Error

ERROR 15 indicates span calibration failure. Possible Causes: β€’ Wrong calibration weight β€’ Unstable platform β€’ Load cell signal issue β€’ Calibration value outside limits Troubleshooting Steps: 1. Use certified calibration weights. 2. Verify entered weight value. 3. Check platform stability. 4. Repeat calibration procedure. 5. Inspect load cell output. If calibration repeatedly fails, inspect the indicator hardware.

Read Fix β†’
ERROR 13

Zero Calibration Error

ERROR 13 indicates failure during zero calibration. Possible Causes: β€’ Incorrect zero calibration procedure β€’ Load cell signal out of range β€’ Mechanical restriction β€’ Damaged load cell Troubleshooting Steps: 1. Remove all weight. 2. Verify platform movement. 3. Inspect load cell signal. 4. Repeat zero calibration. 5. Replace faulty components if required. Always verify calibration after repair.

Read Fix β†’
ERROR 10

Tare Overflow Error

ERROR 10 occurs when the tare value exceeds allowable limits. Possible Causes: β€’ Excessive tare weight β€’ Incorrect tare operation β€’ Configuration issue Troubleshooting Steps: 1. Remove container from platform. 2. Clear tare value. 3. Verify capacity settings. 4. Apply tare again within limits. Avoid exceeding indicator capacity.

Read Fix β†’
ERROR 09

Zero Range Error

ERROR 09 indicates the current zero value is outside the allowable range. Possible Causes: β€’ Weight on platform during startup β€’ Incorrect calibration β€’ Load cell offset β€’ Mechanical obstruction Troubleshooting Steps: 1. Remove all weight. 2. Check for mechanical interference. 3. Restart indicator. 4. Perform zero calibration. 5. Verify load cell output. Replace faulty load cells if zero cannot be restored.

Read Fix β†’
ERROR 08

Unstable Weight Error

ERROR 08 appears when a weighing operation is attempted while the scale is unstable. Possible Causes: β€’ Platform vibration β€’ Air movement β€’ Mechanical instability β€’ Load cell problem Troubleshooting Steps: 1. Wait for stability indicator. 2. Remove vibration sources. 3. Verify proper installation. 4. Inspect load cell mounting. 5. Repeat operation after stabilization. Continuous instability may indicate load cell failure.

Read Fix β†’
ERROR 05

Keypad Error

ERROR 05 indicates a keypad malfunction. Possible Causes: β€’ Stuck key β€’ Damaged keypad membrane β€’ Moisture contamination β€’ Keypad circuit fault Troubleshooting Steps: 1. Check all front panel buttons. 2. Verify no key is stuck. 3. Clean the keypad area. 4. Restart the indicator. 5. Replace damaged keypad if necessary. Persistent keypad faults may require service.

Read Fix β†’
ERROR 02

Load Cell Signal Error

ERROR 02 indicates abnormal load cell input signal. Possible Causes: β€’ Load cell disconnected β€’ Damaged load cell cable β€’ Incorrect wiring β€’ Junction box fault β€’ A/D conversion issue Troubleshooting Steps: 1. Inspect load cell wiring. 2. Verify excitation voltage. 3. Check junction box terminals. 4. Test load cell output signal. 5. Repair or replace damaged wiring. If the signal remains abnormal, inspect the indicator A/D circuit.

Read Fix β†’
ERROR 01

Initialization Failure Error

ERROR 01 occurs when the indicator cannot complete its startup initialization sequence. Possible Causes: β€’ Weight present on platform during startup β€’ Unstable platform β€’ Load cell signal fluctuation β€’ Load cell wiring issue β€’ Internal indicator fault Troubleshooting Steps: 1. Remove all weight from the platform. 2. Switch OFF the indicator. 3. Verify the platform moves freely. 4. Check load cell wiring connections. 5. Restart the indicator. 6. Verify the display returns to normal operation. If ERROR 01 continues, inspect the load cell and indicator electronics.

Read Fix β†’
ADC ERROR

ADC Signal Error

ADC Error indicates the indicator cannot properly read the load cell signal. Possible Causes: β€’ Broken load cell cable β€’ Damaged load cell β€’ Internal ADC circuit fault β€’ Electrical interference Troubleshooting Steps: 1. Inspect load cell cable. 2. Verify excitation voltage. 3. Disconnect and reconnect wiring. 4. Test load cell output. 5. Replace damaged components if necessary. Professional service may be required for ADC circuit repairs.

Read Fix β†’
UNDER ZERO

Under Zero Error

The scale is detecting a negative signal below the allowable zero range. Possible Causes: β€’ Load cell wiring issue β€’ Damaged load cell β€’ Mechanical interference β€’ Incorrect calibration Troubleshooting Steps: 1. Remove all weight from platform. 2. Check load cell wiring. 3. Inspect platform movement. 4. Re-calibrate the scale. 5. Replace faulty load cell if required.

Read Fix β†’
OVER CAPACITY

OVER Capacity Error

The indicator displays OVER when the applied load exceeds the maximum capacity. Possible Causes: β€’ Excessive load on platform β€’ Incorrect capacity settings β€’ Mechanical obstruction Troubleshooting Steps: 1. Remove excess weight. 2. Check scale capacity settings. 3. Inspect platform movement. 4. Verify load cell installation. 5. Restart the indicator. Do not continue using the scale with overload conditions as this may damage the load cell.

Read Fix β†’
ERROR 06

ERROR 06 Calibration Failure

ERROR 06 appears when the calibration process cannot be completed successfully. Possible Causes: β€’ Incorrect calibration weight β€’ Unstable load cell signal β€’ Calibration weight applied incorrectly β€’ Internal hardware fault Troubleshooting Steps: 1. Verify correct calibration weight. 2. Ensure platform is empty before calibration. 3. Repeat calibration procedure. 4. Check load cell wiring. 5. Restart the indicator. 6. Perform calibration again. If the error persists, inspect the load cell and indicator electronics.

Read Fix β†’
ERROR 41

ERROR 41 Load Cell Signal Instability

ERROR 41 indicates unstable load cell signal input. Possible Causes: β€’ Damaged load cell β€’ Loose load cell cable β€’ Moisture inside junction box β€’ ADC circuit instability β€’ Improper grounding Troubleshooting Steps: 1. Inspect load cell wiring. 2. Check excitation voltage. 3. Verify junction box connections. 4. Remove moisture or corrosion. 5. Replace defective load cell if required. 6. Restart the indicator and verify stable readings. If the issue remains after inspection, professional service may be required.

Read Fix β†’