Warm-Up Error occurs when the balance has not reached a stable operating temperature before weighing or calibration.
Possible Causes
1.) Balance switched ON recently.
2.) Temperature changes in the laboratory.
3.) Cold transportation conditions.
4.) Immediate calibration after power ON.
5.) Frequent power interruptions.
Troubleshooting Procedure
1.) Keep the balance switched ON.
2.) Allow sufficient warm-up time as recommended for the model.
3.) Ensure the room temperature is stable.
4.) Verify the balance is level.
5.) Perform calibration after warm-up.
6.) Verify weighing stability.
Prevention
1.) Allow warm-up before critical measurements.
2.) Avoid switching the balance ON and OFF repeatedly.
3.) Maintain a controlled laboratory environment.