Calibration Failed Error indicates that the balance could not complete the calibration process successfully.
Possible Causes
1.) Incorrect calibration weight.
2.) Unstable weighing conditions.
3.) Draft shield left open.
4.) Balance not levelled.
5.) Electrical interference.
6.) Calibration interrupted.
Troubleshooting Procedure
1.) Verify the certified calibration weight.
2.) Level the balance.
3.) Close the draft shield.
4.) Remove vibration sources.
5.) Repeat the calibration procedure.
6.) Verify calibration results.
7.) Contact authorized service if the problem continues.
Prevention
1.) Use certified calibration weights.
2.) Perform calibration in a stable environment.
3.) Avoid interruptions during calibration.