Internal Calibration Error occurs when the balance cannot complete its built-in calibration sequence successfully.
Possible Causes
1.) Balance not levelled.
2.) Unstable installation surface.
3.) Air drafts during calibration.
4.) Internal calibration mechanism obstruction.
5.) Electronic fault.
Troubleshooting Procedure
1.) Verify that the balance is level.
2.) Close the draft shield.
3.) Remove nearby vibration sources.
4.) Allow the balance to warm up.
5.) Restart the internal calibration process.
6.) Verify calibration results with a certified test weight.
Prevention
1.) Keep the balance in a stable environment.
2.) Perform routine maintenance.
3.) Verify calibration regularly.