This error occurs when the weighing pan or load receptor is not installed correctly, preventing accurate measurements.
Possible Causes
1.) Weighing pan not seated correctly.
2.) Foreign object beneath the pan.
3.) Incorrect assembly after cleaning.
4.) Bent or damaged pan support.
5.) Mechanical obstruction.
Troubleshooting Procedure
1.) Remove the weighing pan.
2.) Inspect the pan support.
3.) Remove dust or foreign objects.
4.) Reinstall the weighing pan correctly.
5.) Verify free movement of the pan.
6.) Perform a test weighing.
Prevention
1.) Reinstall the pan carefully after cleaning.
2.) Keep the weighing chamber free from debris.
3.) Inspect the pan support during routine maintenance.