Under Capacity Error indicates that the load cell signal is below the expected operating range.
Possible Causes
1.) Load receptor not installed correctly.
2.) Internal load cell issue.
3.) Mechanical obstruction.
4.) Improper assembly after cleaning.
5.) Internal hardware fault.
Troubleshooting Procedure
1.) Inspect the weighing pan installation.
2.) Check for foreign objects beneath the pan.
3.) Restart the balance.
4.) Verify zero indication.
5.) Perform calibration verification.
6.) Contact authorized service if the error remains.
Prevention
1.) Install the weighing pan correctly.
2.) Keep the weighing chamber clean.
3.) Avoid applying upward force to the weighing pan.
Is your scale beyond repair?