Essae DS-215N Span Calibration Procedure
Span calibration adjusts weighing accuracy across the operating capacity range.

Procedure
1.) Complete zero calibration first.
2.) Wait for CAL SP display.
3.) Place approved calibration weight.
4.) Verify correct weight value.
5.) Start span calibration.
6.) Wait for calibration completion.
7.) Observe span count display.
8.) Remove test weights.
9.) Verify weighing accuracy.
10.) Save calibration data.

Recommended Calibration Weights
1.) 30kg Scale → 20kg Test Weight
2.) 60kg Scale → 40kg Test Weight
3.) 100kg Scale → 60kg Test Weight
4.) 150kg Scale → 100kg Test Weight
5.) 300kg Scale → 200kg Test Weight

Benefits
1.) Improved weighing accuracy.
2.) Better repeatability.
3.) Compliance with quality systems.

Looking for a new weighing machine?

Upgrade your operations with our latest high-precision scales. We offer a wide catalog of reliable and certified machines.

Related Calibration Guides

Essae SI-810 Complete Software Administration Guide

Essae SI-810 Complete Software Administration Guide

ESSAE

View Guide
Essae SI-810 Software FAQ Guide

Essae SI-810 Software FAQ Guide

ESSAE

View Guide
Essae SI-810 Software Best Practices Guide

Essae SI-810 Software Best Practices Guide

ESSAE

View Guide

Related Error Codes

Essae SI-810 Database Corruption Error

DATABASE CORRUPTION

View Error

Essae SI-810 File Access Error

FILE ACCESS ERROR

View Error

Essae SI-810 Printer Connection Error

PRINTER CONNECTION ERROR

View Error

Related Technical Manuals

Essae AX Series PDF Manual Download

Download PDF

Essae AX Series Technical Reference Manual

Download PDF

Essae AX Series Laboratory Balance Manual PDF

Download PDF

Essae AX Series Calibration and Maintenance Manual

Download PDF